Visual performance of the Tecnis one-piece lens ZCB00 (2024)

  • Journal List
  • Clin Ophthalmol
  • v.5; 2011
  • PMC3258090

As a library, NLM provides access to scientific literature. Inclusion in an NLM database does not imply endorsem*nt of, or agreement with, the contents by NLM or the National Institutes of Health.
Learn more: PMC Disclaimer | PMC Copyright Notice

Visual performance of the Tecnis one-piece lens ZCB00 (1)

Dove Medical PressThis ArticleSubscribeSubmit a ManuscriptSearchFollowDovepressClinical Ophthalmology

Clin Ophthalmol. 2011; 5: 1803–1808.

Published online 2011 Dec 23. doi:10.2147/OPTH.S27324

PMCID: PMC3258090

PMID: 22267915

Sherein S Wahba, Riad F Riad, Fathy F Morkos, Abdallah K Hassouna, and Maged M Roshdy

Author information Copyright and License information PMC Disclaimer

Abstract

Synopsis

The aspheric, one-piece design of the Tecnis® ZCB00 intraocular lens provided higher mesopic contrast sensitivity and relatively constant refraction at different pupil sizes compared to the Sensar® AR40e.

Purpose

Comparing the impact on functional vision of the aspheric Tecnis® ZCB00 one-piece lens compared to the spherical Sensar® AR40e three-piece lens.

Setting

National Eye Hospital, Cairo, Egypt.

Methods

In this prospective study, we implanted 32 Tecnis ZCB00 (Abbott Medical Optics, Santa Ana, CA) and 30 Sensar AR40e (Pharamacia and Upjohn LLC, North Peapack, NJ) lenses. Twenty-three patients in the ZCB00 group and 13 patients in the AR40e group provided reliable, reproducible data, and were therefore included in the statistical analysis. After 8 weeks, we measured photopic contrast sensitivity (PCS) at 85 cd/m2 and mesopic contrast sensitivity (MCS) at 5 cd/m2. Uncorrected and best-corrected visual acuity were measured. Myopic shift was analyzed by comparing the manifest refraction spherical equivalent at 3 days and 8 weeks after surgery in the normal and dilated pupil.

Results

The Tecnis ZCB00 group showed significantly better MCS than the Sensar AR40e group at a spatial frequency of six cycles per degree (cpd) (P = 0.037), but not at 3, 12, or 18 cpd (P = 0.299, P = 0.226, and P = 0.396, respectively). There were no significant differences between groups in corrected distance visual acuity (P = 0.175) or PCS at 3, 6, 12, or 18 cpd spatial frequencies (P = 0.440, P = 0.176, P = 0.365, and P = 0.251, respectively). The ZCB00 group showed less myopic shift in normal and dilated pupils between 3 days and 8 weeks after surgery compared to the AR40e group, but this difference was not significant.

Conclusion

The Tecnis one-piece ZCB00 provided higher MCS at moderate spatial frequency than the three-piece Sensar AR40e, and was associated with relatively constant refraction at different pupil sizes.

Keywords: Tecnis ZCB00, Sensar AR40e, contrast sensitivity, myopic shift

Introduction

The decreases in visual acuity and contrast sensitivity that occur with age are usually attributed to changes in the lens.1 The crystalline lens undergoes positive spherical aberration (SA) with age, which disrupts the wave front balance with the cornea and increases the positive SA of the whole eye.2 In young subjects, the crystalline lens compensates for any positive SA in the cornea with negative SA, which results in a low level of SA for the entire eye.3 Traditional intraocular lenses (IOLs) also have positive SA, and therefore, they produce a pseudophakic eye that is no better than an aged eye with a transparent lens in this regard.4

Aspheric IOLs are designed to introduce optical path differences that bend the light rays to a single focus on the retina. There are several IOL designs that compensate for the SA by making one or both surfaces aspheric.5

Several studies610 have suggested that implantation of these aspheric IOLs might improve contrast sensitivity and enhance patient quality of life. However, others11 found no significant differences in contrast sensitivity between aspheric and spherical IOLs.

In this study, we compared the performance of the Tecnis® ZCB00 (Abbott Medical Optics, Santa Ana, CA) and Sensar® AR40e (Pharamacia and Upjohn LLC, North Peapack, NJ) with regard to their contribution to visual outcome and contrast sensitivity. Tecnis ZCB00 is a one-piece 6.0 mm biconvex hydrophobic acrylic lens with anterior aspheric surface that resulted in a negative SA of −0.27 μm and frosted continuous 360° posterior square edge. Sensar AR40e is a three-piece 6 mm biconvex spheric hydrophobic acrylic lens with OptiEdge™ (Abbott Medical Optics) design and polymethylmethacrylate (PMMA) haptic with 5° angulation (Table 1).

Table 1

Features of the two IOL types used in this study

Tecnis® ZCB00Sensar® AR40e
One-piece IOLThree-piece IOL
6.0-mm biconvex, anterior aspheric surface, square optic edge6-mm biconvex spheric hydrophobic acrylic optic
Frosted continuous 360° posterior square edgeOptiEdge® design
Haptic offset from optic5° haptic angulation
UV blocking hydrophobic acrylic hapticPMMA haptic
13 mm D, RI: 1.4713 mm D, RI: 1.47

Open in a separate window

Abbreviations: IOL, intraocular lens; PMMA, polymethylmethacrylate; D, diameter; RI, refractive index.

Patients and methods

A randomized prospective study was conducted in 62 eyes in patients that required cataract surgery at the National Eye Hospital. Exclusion criteria included ocular surface pathology, uncontrolled glaucoma, existing retinal or optic disc pathology, and systemic diseases that could potentially affect the vision, like diabetes. After cataract removal, the patients were divided into two groups. The first group of 32 eyes received implants of aspheric Tecnis ZCB00 lenses (Abbott Medical Optics) and the second group of 30 eyes received implants of spherical Sensar AR40e lenses (Abbott Medical Optics). Informed consent was obtained from all participants and the study was reviewed by the National Eye Hospital ethics committee.

Surgical procedure

All surgical procedures were performed with the same technique, and the same phaco machine. After administration of periocular anesthesia, a corneal incision of 2.4 mm was performed in the steep meridian when possible, a 5-mm continuous capsulorhexis was performed, followed by phacoemulsification cataract extraction. Next, an IOL was implanted with an injector; then, the IOL was centered and the meticulous ophthalmic viscoelastic device was washed.

Evaluation

All patients had a full preoperative ocular examination, including refraction testing, best corrected distance visual acuity (CDVA) testing, tonometry, funduscopy, slit-lamp examination, and corneal topography. Patients were examined directly after surgery for the baseline data; then, again, at 2–3 days, 2 weeks, and 6–8 weeks after surgery. The data presented included refraction, CDVA (recorded in logarithm of minimum angle of resolution [logMAR]), corneal topography, photopic contrast sensitivity (PCS), and mesopic contrast sensitivity (MCS).

Contrast sensitivity was measured with the CSV1000E test (Good-Lite, Elgin, IL), using sine-wave grating charts to measure spatial frequencies of 3, 6, 12, and 18 cycles per degree (cpd). The CSV1000E test has been used in several studies and has been shown to give reliable, repeatable contrast sensitivity function scores.12 All measurements were performed under photopic conditions at 85 cd/m2 and under mesopic conditions at 5 cd/m2.

According to Wachler and Krueger13 contrast sensitivity values were normalized using the ratio between a patient’s result and the normal score for the patient’s age group. Scores higher or lower than 1.0 indicated contrast sensitivities higher or lower than expected.

Corneal topography was measured with a WaveLight® Allegro Oculyzer with software version 1.16r12 (WAVELIGHT AG, Erlangen, Germany) to ensure corneal regularity.

Statistical analysis

Data were collected, verified, and the changes were calculated with Microsoft Excel 2007. Statistical analyses were performed with MedCalc (v 11.1.1.0, MedCalc Software, Mariakerke, Belgium). The following tests were performed: calculation of the mean, standard deviation (SD), the t-test for independent samples of parametric data with equal variances, and the Welch test for independent samples of parametric data with unequal variances.

Results

Twenty-three patients in the Tecnis ZCB00 group and 13 patients in the SensarAR40e group attended all the follow-up visits in the study. Their data were reliable and reproducible (no same visit variability); therefore, all were included in the statistical analysis.

Patients

The mean age of the patients was 63.8 ± 10.4 years. The male to female ratio was 23:39. The right to left ratio was 20:16. The demographic characteristics of the study groups are reported in Table 2. The mean IOL power was 19.3 ± 5.3 in the Tecnis ZCB00 group, and 17.3 ± 6.3 in the Sensar AR40e group (P = 0.194) (Table 3).

Table 2

Study group demographics and IOL information

Tecnis® ZCB00Sensar® AR40eP value
Mean age (year) ± SD63.6 ± 8.061.6 ± 12.70.354
Male/female8/154/90.065
Right/Left13/107/60.846

Open in a separate window

Abbreviations: IOL, intraocular lens; SD, standard deviation.

Table 3

Myopic shifts at different pupil sizes and mean IOL powers in the two study groups

Myopic shift, undilated pupilMyopic shift, dilated pupilMean IOL power (diopter)
Mean ± SD (diopter)P valueMean ± SD (diopter)P valueMean ± SD (diopter)P value
Tecnis® ZCB00−0.06 ± 0.590.1940.06 ± 0.480.19619.1 ± 5.90.19
Sensar® AR40e−0.32 ± 0.45−0.16 ± 0.3618.7 ± 6.3

Open in a separate window

Abbreviations: IOL, intraocular lens; SD, standard deviation.

Visual acuity

Two months post-operatively, the mean CDVA was 0.89 ± 0.17 in the Tecnis ZCB00 group, and 0.96 ± 0.14 in the Sensar AR40e group, with no significant difference between groups (P = 0.175) (Figure 1).

Open in a separate window

Figure 1

A comparison of mean best-corrected visual acuities measured at 2 months post-operatively.

Contrast sensitivity

The Tecnis ZCB00 one-piece showed better PCS than the Sensar AR40e three-piece at all spatial frequencies (Figure 2, Table 4), but there were no statistically significant differences between groups (P = 0.440, P = 0.176, P = 0.365, and P = 0.251, respectively).

Open in a separate window

Figure 2

Mean photopic contrast sensitivities of the two study groups.

Table 4

Mean photopic contrast sensitivities of the two study groups

Photobic contrast sensitivity (mean)3 cpd6 cpd12 cpd18 cpd
Tecnis® ZCB001.41.50.990.51
Sensar® AR40e1.311.360.870.34

Open in a separate window

Abbreviation: cpd, cycles per degree.

At a spatial frequency of 6 cpd, the MCS was significantly better in the Tecnis ZCB00 one-piece group than in the SensarAR40e three-piece group (P = 0.037; Figure 3, Table 5); however, at spatial frequencies of 3, 12, or 18 cpd, the MCS was not significantly different between the two groups (P = 0.299, P = 0.226, and P = 0.396, respectively).

Open in a separate window

Figure 3

Mean mesopic contrast sensitivities of the two study groups.

Note: **P < 0.05.

Table 5

Mean mesopic contrast sensitivities of the two study groups

Mesopic contrast sensitivity (mean)3 cpd6 cpd12 cpd18 cpd
Tecnis® ZCB001.161.180.670.1
Sensar® AR40e1.290.90.530.19

Open in a separate window

Abbreviation: cpd, cycles per degree.

Myopic shift

In undilated eyes, the change in spherical equivalent from 3 days to 2 months was lower in the Tecnis ZCB00 lenses (−0.06 D) than in Sensar AR40e lenses (−0.32 D), but this difference was not statistically significant (P = 0.194; Table 3). In dilated eyes, the change in spherical equivalent from 3 days to 2 months was lower in the Tecnis ZCB00 lenses (0.06 D) than in the Sensar AR40e lenses (−0.16 D), but this difference was not statistically significant (P = 0.196; Table 3). This indicated that there was less myopic shift with the Tecnis ZCB00 one-piece lenses than with the Sensar AR40e three-piece lenses in both dilated and undilated eyes.

Discussion

Optical aberrations during cataract surgery can be reduced by implanting aspheric IOL. Standard IOLs add positive SA to that already induced by the cornea and, thereby, they decrease optical quality. The goal of the current study was to evaluate the visual performance of aspheric Tecnis ZCB00 one-piece hydrophobic acrylic lenses and compare them to spherical Sensar AR40e three-piece hydrophobic acrylic lenses.

In this study, we found no significant difference between groups in the post-operative CDVA (P = 0.175). In contrast, Bellucci et al10 compared the CDVA (logMAR) in patients that received implants of Tecnis Z9000 or conventional AcrySof SA60AT IOL. They reported a difference of approximately three letters in favor of the Tecnis Z9000 IOL implants. On the other hand, Thiagarajan et al14 found a statistically significant difference in CDVA but not clinically important (relative difference 1.5%), between Akreos Adapt Advanced Optics (AO) (Bausch and Lomb, Inc, Rochester, NY) aspheric intraocular lens (IOL) and the Sensar AR40e spherical acrylic IOL, in favor of the spherical IOL.

In a study by Munoz et al2, there were no statistically significant differences in CDVA among patients that received implants of aspheric Tecnis Z9000 IOL, the AR40e IOL, or the Stabibag (Carl Zeiss Meditec, Inc, Dublin, CA) IOLs. Pandita et al15 evaluated the CDVA in patients that had implants of AcrySof SA60AT, AcrySof Natural SN60AT, or AcrySof IQ SN60WF IOLs; the results were similar among the three groups. Also, Morales et al16 compared visual acuity in patients who had a spherical IOL (SoFlex SE, Bausch and Lomb) implanted in one eye and an aspheric IOL (SofPort AO, Bausch and Lomb) implanted in the fellow eye and there were no statistical differences among the two groups. Thus, the previous studies showed results similar to our results for the CDVA.

In the current study, we found no significant differences in the PCS between the Tecnis ZCB00 Aspheric IOL and the Sensar AR40e spherical IOL at all spatial frequencies. However, at 6 cpd, the MCS in the Tecnis ZCB00 was significantly better than that in the Sensar AR40e. Our results were comparable to those from the study by Denoyer et al17 who compared aspheric Tecnis Z9000 and spherical CeeOn™ Edge 911 (Pharmacia, Kalamazoo, MI) IOLs. That study revealed that the Tecnis Z9000 group showed significantly better MCS at high spatial frequencies; however, the contrast sensitivity under photopic and glare conditions did not differ between the two groups. Also, the study by Morales et al16 was comparable to our results; their study compared aspheric SofPort AO® IOL (Bausch and Lomb) and spherical SoFlex SE® IOL (Bausch and Lomb). They found no statistical difference between groups under photopic conditions but better performance was observed with the aspheric IOL at low spatial frequencies under mesopic conditions.

Ohtani et al18 found no significant differences in contrast sensitivity, under either photopic or mesopic conditions, between patients that received implants of Tecnis Z9003 IOLs or the AR40e. Munoz et al2 found that both the PCS and MCS were similar in patients that received implants of Tecnis Z9000, Sensar AR40e, or Stabibag IOLs. Also, Thiagarajan et al14 found no statistically significant differences in contrast sensitivity between Akreos Adapt Advanced Optics aspheric intraocular lens (IOL) and the Sensar AR40e spherical acrylic IOL.

However, our results contrasted with those of Pandita et al.15 They found that, compared to the AcrySof SA60AT and AcrySof SN60AT, the AcrySof IQ IOLs provided significantly higher PCS at 18 cpd, higher MCS with a 4.0 mm aperture at all spatial frequencies without glare and with glare, and higher MCS with a 6.0 mm aperture at all spatial frequencies without glare and with glare. Also, Bellucci et al10 showed that, compared to AcrySof SA60AT IOLs, the Tecnis Z9000 IOLs provided better contrast sensitivity at 3, 6, 12, and 18 cpd in both photopic and mesopic conditions, with a peak improvement at 18 cpd.

There are several possible explanations for the lack of significant differences in visual performance among different IOLs. First, both IOLs used in our study were made of the same hydrophobic acrylic material. Second, the aspheric nature of the Tecnis one-piece ZCB00 will improve the SA of the eye, but this effect will mainly appear in mesopic conditions, as was the case in our study. Third, an improvement in the optical quality of the eye may not necessarily lead to an improvement in the quality of vision.

In the current study, there was less myopic shift with the Tecnis ZCB00 than with the Sensar AR40e in different pupil sizes, but this difference was not statistically significant. Our results were comparable to those of Bellucci et al19 who found that, in contrast to the myopic shift that tends to occur with the implantation of a spherical IOL, no shift occurred when they implanted the Tecnis one-piece IOL.

Nejima et al20 found that, although the MA60AC three-piece IOL showed significant forward shift and myopic refractive changes after surgery, the SA60AT one-piece IOL displayed little axial movement and was associated with highly stable refraction after surgery. Szurman et al21 evaluated the aspheric design of the Tecnis one-piece and how it correlated to patient benefits in 102 patients; they found no myopic shifts at larger diameters. Petermeier et al22 evaluated the influence of the pupil size on clinical results and objective parameters of optical quality of the Tecnis one-piece (ZCB00) IOLs. His study included 51 eyes; they found the defocus curves with a 3-mm and a 5-mm pinhole were not statistically significantly different, which was comparable to our results. Taken together, these results suggested that the one-piece design conferred more stable refraction at different pupil sizes.

In conclusion, we found that the aspheric property of the Tecnis one-piece ZCB00 IOL provided higher MCS at moderate spatial frequency than the spherical three-piece Sensar AR40e. Furthermore, its one-piece property was associated with a relatively constant refraction at different pupil sizes. In the future, a meta-analysis might provide conclusive evidence that could confirm or negate the real benefit of one IOL model compared to the others.

Footnotes

Disclosure

Sherein S Wahba and Maged M Roshdy received travel support from Abbott Medical Optics, Inc, Santa Ana, CA. Riad F Riad, Fathy F Morkos and Abdallah K Hassouna received travel support and equipment from Abbott Medical Optics, Inc, Santa Ana, CA and Alcon Laboratories, Inc, CA. The authors have no other financial interest in the products mentioned in this paper. This paper was presented at the XXVIII Congress of the ESCRS, Paris, France, September 4–8, 2010.

References

1. Owsley C, Sekuler R, Siemsen D. Contrast sensitivity throughout adulthood. Vision Res. 1983;23(7):689–699. [PubMed] [Google Scholar]

2. Muñoz G, Albarrán-Diego C, Montés-Mico R, Rodríguez-Galietero A, Alió JL. Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens. J Cataract Refract Surg. 2006;32(8):1320–1327. [PubMed] [Google Scholar]

3. Artal P, Berrio E, Guirao A, Piers P. Contribution of the cornea and internal surfaces to the changes of ocular aberrations with age. J Opt Soc Am A Opt Image Sci Vis. 2002;19(1):137–143. [PubMed] [Google Scholar]

4. Atchison DA. Design of aspheric intraocular lenses. Ophthalmic Physiol Opt. 1991;11(2):137–146. [PubMed] [Google Scholar]

5. Franchini A. Compromise between spherical and chromatic aberration and depth of focus in aspheric intraocular lenses. J Cataract Refract Surg. 2007;33(3):497–509. [PubMed] [Google Scholar]

6. Packer M, Fine IH, Hoffman RS, Piers PA. Prospective randomized trial of an anterior surface modified prolate intraocular lens. J Refract Surg. 2002;18(6):692–696. [PubMed] [Google Scholar]

7. Mester U, Dillinger P, Anterist N. Impact of a modified optic design on visual function: clinical comparative study. J Cataract Refract Surg. 2003;29(4):652–660. [PubMed] [Google Scholar]

8. Kershner RM. Retinal image contrast and functional visual performance with aspheric, silicone, and acrylic intraocular lenses. Prospective evaluation. J Cataract Refract Surg. 2003;29(9):1684–1694. [PubMed] [Google Scholar]

9. Packer M, Fine IH, Hoffman RS, Piers PA. Improved functional vision with a modified prolate intraocular lens. J Cataract Refract Surg. 2004;30(5):986–992. [PubMed] [Google Scholar]

10. Bellucci R, Scialdone A, Buratto L, et al. Visual acuity and contrast sensitivity comparison between Tecnis and AcrySof SA60AT intraocular lenses: A multicenter randomized study. J Cataract Refract Surg. 2005;31(4):712–717. Erratum in: J Cataract Refract Surg. 2005;31(10):1857. [PubMed] [Google Scholar]

11. Kasper T, Bühren J, Kohnen T. Visual performance of aspherical and spherical intraocular lenses: intraindividual comparison of visual acuity, contrast sensitivity, and higher-order aberrations. J Cataract Refract Surg. 2006;32(12):2022–2029. [PubMed] [Google Scholar]

12. Quesnel NM, Lovasik JV, Ferremi C, Boileau M, Ieraci C. Laser in situ keratomileusis for myopia and the contrast sensitivity function. J Cataract Refract Surg. 2004;30:1209–1218. [PubMed] [Google Scholar]

13. Wachler BS, Krueger RR. Normalized contrast sensitivity values. J Refract Surg. 1998;14(4):463–466. [PubMed] [Google Scholar]

14. Thiagarajan M, McClenaghan R, Anderson DF. Comparison of visual performance with an aspheric intraocular lens and a spherical intraocular lens. J Cataract Refract Surg. 2011;37(11):1993–2000. [PubMed] [Google Scholar]

15. Pandita D, Raj SM, Vasavada VA, Vasavada VA, Kazi NS, Vasavada AR. Contrast sensitivity and glare disability after implantation of AcrySof IQ Natural aspherical intraocular lens: prospective randomized masked clinical trial. J Cataract Refract Surg. 2007;33(4):603–610. [PubMed] [Google Scholar]

16. Morales EL, Rocha KM, Chalita MR, Nosé W, Avila MP. Comparison of optical aberrations and contrast sensitivity between aspheric and spherical intraocular lenses. J Refract Surg. 2011;27(10):723–728. [PubMed] [Google Scholar]

17. Denoyer A, Le Lez ML, Majzoub S, Pisella PJ. Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation: effect of contrast sensitivity and wavefront aberration improvements on the quality of daily vision. J Cataract Refract Surg. 2007;33(2):210–216. [PubMed] [Google Scholar]

18. Ohtani S, Miyata K, Samejima T, Honbou M, Oshika T. Intraindividual Comparison of Aspherical and Spherical Intraocular Lenses of Same Material and Platform. Ophthalmology. 2009;116:896–901. [PubMed] [Google Scholar]

19. Bellucci R, Morselli S, Piers P. Comparison of wavefront aberrations and optical quality of eyes implanted with five different intraocular lenses. J Refract Surg. 2004;20(4):297–306. [PubMed] [Google Scholar]

20. Nejima R, Miyai T, Kataoka Y, et al. Prospective intrapatient comparison of 6.0-millimeter optic single-piece and 3-piece hydrophobic acrylic foldable intraocular lenses. Ophthalmology. 2006;113(4):585–590. [PubMed] [Google Scholar]

21. Szurman P, Petermeier K. Aspheric IOLs benefit quality of vision. Clinical results after Tecnis 1-Piece IOL implantation are promising. Cataract & Refractive Surgery Today Europe. 2009:17–18. [Google Scholar]

22. Petermeier K, Frank C, Gekeler F, Spitzer MS, Messias A, Szurman P. Influence of the pupil size on visual quality and spherical aberration after implantation of the Tecnis 1-piece intraocular lens. Br J Ophthalmol. 2011;95(1):42–45. [PubMed] [Google Scholar]

Articles from Clinical Ophthalmology (Auckland, N.Z.) are provided here courtesy of Dove Press

Visual performance of the Tecnis one-piece lens ZCB00 (2024)

FAQs

Is the ZCB00 lens good? ›

The Tecnis ZCB00 group showed significantly better MCS than the Sensar AR40e group at a spatial frequency of six cycles per degree (cpd) (P = 0.037), but not at 3, 12, or 18 cpd (P = 0.299, P = 0.226, and P = 0.396, respectively).

What type of lens is ZCB00? ›

TECNIS Monofocal 1-Piece IOL (ZCB00)
TECNIS Monofocal 1-Piece IOL (ZCB00)OPTIC CHARACTERISTICS
Diameter:6.0 mm
Shape:Biconvex, anterior aspheric surface, square optic edge
Material:UV-blocking hydrophobic acrylic
Refractive Index:1.47 at 35° C
2 more rows

Which is better ICB00 or ZCB00? ›

The mean defocus curve in ICB00 and ZCB00 were similar at 1D and found to be nonsignificant (P = 0.914). ICB00 has significantly better visual acuity across defocus levels from −0.50 to −2.50 D (P < 0.01).

What is the most popular lens after cataract surgery? ›

Monofocal IOLs

The most common type of lens used with cataract surgery is called a monofocal IOL. It has one focusing distance. It is set to focus for up close, medium range or distance vision. Most people have them set for clear distance vision.

What is the best lens for cataract surgery with astigmatism? ›

If you want to avoid wearing distance glasses after cataract surgery and have astigmatism, a toric lens might be appropriate. A presbyopia-correcting lens may be best if you want to avoid wearing distance and reading glasses after cataract surgery.

Which cataract lens is best for driving? ›

A monofocal lens is also referred to as a single-focus IOL. It provides the sharpest focus at one particular distance, usually from afar which is helpful with driving and other activities where distance vision is required.

What is the price of Tecnis lens? ›

The most popular Johnson and Johnson monofocal lens is TECNIS Monofocal whose price is around Rs. 45,000 to Rs. 55,000.

What is the cost of Johnson and Johnson cataract lens? ›

On average, Johnson & Johnson cataract lenses can range between 45,000 to 90,000 rupees. However, it's always a good idea to consult with your ophthalmologist or eye care center for the most accurate pricing based on your specific needs.

Is Tecnis lens monofocal? ›

Monovision Excellence: As a monovision lens, Tecnis Monofocal provides clarity for either distance or near, based on the individual's preference and lifestyle needs. This adaptability makes it a favored choice among many.

Which IOL is best for distance vision? ›

The most popular intraocular lenses are monofocal. Monofocal lenses sharpen only one focus (far, intermediate, or near), and do not correct astigmatism. Most people use monofocal IOLs to clarify distance vision.

Which lens is better monofocal or multifocal? ›

While monofocal lenses provide focus at one distance only, multifocal lenses allow for vision at a range of distances, meaning they can correct both near, intermediate, and far distances. These are a more popular option for those who want to also eliminate their dependency on glasses.

What is the difference between tecnis and eyhance? ›

Compared to its traditional monofocal analog, the TECNIS 1-piece IOL ZCB00, the Eyhance IOL has a modified aspheric anterior surface that facilitates a steady increase in lens power within the central 1 mm diameter of the IOL optic.

What is the newest lens for cataract surgery in 2024? ›

The Light Adjustable Lens (LAL) by RxSight is a newer IOL option that enables patients to trial and customize their vision after cataract surgery, taking customized IOL powers to the next level. LAL uses a photosensitive material that adjusts lens power when it is exposed to UV light.

Are the more expensive cataract lenses worth it? ›

See Clearly Without Glasses With Premium Lenses

You will be able to see things far away clearly, but you will need to wear glasses for near activities, such as reading, writing, or sewing. The main reason that most people upgrade to premium lenses is for greater flexibility with their vision in their day-to-day lives.

Is it better to be nearsighted or farsighted after cataract surgery? ›

If your daily lifestyle includes a lot of outdoor activities, such as driving or watching TV, then farsightedness after surgery might be a great fit. You'll have clear distance vision for the things you do most, and close-up tasks can be easily tackled with a well-chosen pair of reading glasses.

Which eye lens is best? ›

Progressive lenses have clear benefits compared to bifocal and trifocal lenses. Progressive lenses have a power for the near, intermediate and distance zones – and all of this in a lens with a seamless transition and no dividing line. The power is smooth from top to bottom and moves from distance to near vision.

What is the best intraocular lens for macular degeneration? ›

Clinical results in AMD patients have been described for a total of seven types of IOLs recommended for AMD: an implantable miniature telescope (IMT), IOL-VIP System, Lipsh*tz macular implant (LMI), sulcus-implanted Lipsh*tz macular implant, LMI-SI, Fresnel Prism Intraocular Lens, iolAMD and Scharioth Macula Lens.

What is the best trifocal lens for cataract surgery? ›

The Only FDA-Approved Trifocal

That is the PanOptix trifocal, IOL. The other IOLs available do not match the performance of this new technology. The PanOptix trifocal IOL is the best of both worlds. It provides crisp, clear monofocal-like vision but at all three refractive distances.

What are the disadvantages of Eyhance? ›

Premium lenses can help you see very well, and many can reduce or eliminate your need for reading glasses altogether. But premium lenses can cause glare and halos due to imperfections in the lens.

Top Articles
Woo Pig Softball Tournament 2023
Nc Alvin
Huggies Size 4 Walgreens
Benchmark Physical Therapy Jobs
Greet In Cheshire Crossword Clue
Td Share The Green Referral Credit
Boomerang Uk Screen Bug
Coverwood Terriers For Sale
Autozone Locations Near Me
The Ultimate Guide To Jelly Bean Brain Leaks: Causes, Symptoms, And Solutions
Wow Rock Wall Locked
Creepshot. Org
Lonely Ghost Discount Codes - 20% Off | September 2024
Po Box 6726 Portland Or 97228
Lecture Tutorials For Introductory Astronomy Answer Guide
Black Friday 2024, Black Friday 2025 and further
Roses Gordon Highway
Crazy 8S Cool Math
Banette Gen 3 Learnset
Nwi Police Blotter
352-730-1982
13.2 The F Distribution and the F Ratio - Statistics | OpenStax
Truecarcin
Optum Primary Care - Winter Park Aloma
How Much Is Cvs Sports Physical
Bbc Weather Boca Raton
Camwhor*s Bypass 2022
Kickflip Seeds
Sold 4 U Hallie North
What Does Spd2 Mean On Whirlpool Microwave
Nehemiah 6 Kjv
Kobe Express Bayside Lakes Photos
Susan Dey Today: A Look At The Iconic Actress And Her Legacy
Banette Gen 3 Learnset
Academy Sports Meridian Ms
Phasmophobia Do As I Command Challenge
Ontpress Fresh Updates
Power Outage Map National Grid
Proto Ultima Exoplating
Vogler Funeral Home At Forsyth Memorial Park
Arti kata petang-petang - Kamus Besar Bahasa Indonesia (KBBI) Online
Netdania.com Gold
Wells Fargo Arena Des Moines Seating Chart Virtual View
Henkels And Mccoy Pay Stub Portal
Feetfinder Reviews Trustpilot
What is Landshark Beer?
1636 Fire Red Squirrels
Motorsports Nation | Powersports Dealer in Waterford, CT
The most memorable songs from '90s movies
Samanthaschwartz Fapello
Tetris Google Sites
Cb2 South Coast Plaza
Latest Posts
Article information

Author: Golda Nolan II

Last Updated:

Views: 5381

Rating: 4.8 / 5 (78 voted)

Reviews: 93% of readers found this page helpful

Author information

Name: Golda Nolan II

Birthday: 1998-05-14

Address: Suite 369 9754 Roberts Pines, West Benitaburgh, NM 69180-7958

Phone: +522993866487

Job: Sales Executive

Hobby: Worldbuilding, Shopping, Quilting, Cooking, Homebrewing, Leather crafting, Pet

Introduction: My name is Golda Nolan II, I am a thoughtful, clever, cute, jolly, brave, powerful, splendid person who loves writing and wants to share my knowledge and understanding with you.